Rabu, 09 Maret 2016

LENTON Laboratory Ovens - Welland range - 300°C

A modern range of ovens, with a maximum operating temperature of 300°C, providing a combination of excellent performance and reliability.

Increased power and low thermal mass encased fibre insulation ensure both fast heat up times and reduced recovery times. 

Reduced holding power once at set temperature, together with the insulation, makes the range economical. Cool safe outer case temperature.

Both gravity and forced air circulation models are available with a wide choice of control options, allowing the most critical performance criteria to be met. Where processes involve the liberation of flammable vapours, a stoving and curing option is available and for processes involving large amounts of water, a moisture extraction option is available.
 
 Welland laboratory ovens
Ovens without fansOvens with fans
ModelWN 30WN 60WN 120WN 200WF 30WF 60WF 120WF 200
Maximum temperature (°C)300300300300300300300300
Chamber dimensions 
(mm - h / w / d)
250
330
320
350
392
420
450
492
520
700
590
520
300
292
320
400
392
420
500
492
520
750
792
520
External dimensions
(mm - h / w / d)
470
655
470
570
765
570
670
865
670
920
965
670
470
665
470
570
765
570
670
865
670
920
965
670
Chamber capacity (l)27581152152866128230
Weight (kg)3045607530456075
ShelvesNumber supplied
Maximum possible
Maximum load / shelf (kg)
Maximum load (kg)

2
3
10
20

2
5
10
20

2
9
10
40

2
15
10
50

2
3
10
20

2
5
10
30

2
9
10
40

2
15
10
50
Power rating at 240V (watts)750100015002250100015002002700
Holding power at maximum temperature  (watts)30048072011603506008001250
Temperature uniformity
(at maximum temperature as a %)
+2.3+2.3+2.7+3.5+1.0+1.0+1.0+1.0
Temperature stability with PID control (°C)+0.5+0.5+0.5+0.5+0.2+0.2+0.2+0.2
Heat up time (minutes) 240V to:                  100°C
                                                                            200°C
                                                                            300°C
12
26
52
12
26
52
12
26
52
14
29
58
4.5
12
25
4.5
12
25
4.5
12
25
5.5
14
30
Recovery time (minutes) with door              100°C
open for 60 seconds to:                                  200°C
                                                                             300°C
2.5
5
8.5
2.5
5
8.5
2.5
5
8.5
3
6
10
1
2.5
4
1
2.5
4
1
2.5
4
1.5
3
5
Air exchanges (volume - l / hr at 100°C)
- with standard fan
- with exhaust fan

N/A
N/A

N/A
N/A

N/A
N/A

N/A
N/A

1400
10,000

1400
10,000

1400
10,000

1400
10,000
Air exchanges / hour
- with standard fan
- with optional exhaust fan

N/A
N/A

N/A
N/A

N/A
N/A

N/A
N/A

50
360

21
153

11
79

6
44

Tidak ada komentar: